LKO-1500

1 / 4
LKO-1500
Discontinued Product
Discontinued Product
Index: WMGBLKO1500
Receiver
Description
Technical Specification
Files
  • Description

    LKO-1500 receiver is facilitates the work of the user to with many features, including visualization of the cable routing in the tested object or a greater number of frequencies to choose from. It is also distinguished by the option of tracing cable short-to-earth faults with the use of additional accessories.

    LKO-1500 receiver

    • Active mode:
      • work with LKN-1500 transmitter at the frequencies of 273 Hz, 526 Hz, 1024 Hz, 8928 Hz, 33 kHz,
      • work with transmitters other than LKN-1500 at the frequencies of 491 Hz, 512 Hz, 982 Hz, 2000 Hz, 2048, Hz 8440 Hz, 9828 Hz, 10000 Hz.
    • Passive mode – work at the frequencies of:
      • 50 Hz, 60 Hz, 100 Hz, 300 Hz, 550 Hz, 1450 Hz,
      • RADIO (10...36 kHz),
      • SB (48 Hz...14 kHz).
    • Location depth up to 10 m.
    • Operation with external probes.
  • Technical Specification

    Technical specifications


    Parameter Description
    Ingress protection IP54
    Power supply Ni-MH 6 V/2000 mAh battery pack, 5x AA 1.5 V battery, 5x Ni-MH 1.2 V rechargeable battery
    Operating temperature -20…+55°C
    Storage temperature -50…+70°C
    Nominal temperature 23 ± 2°C
    Relative humidity no more than 95% at the temperature of +30°С

    The main specifications


    No Parameter Value Value
    1. Rated values of operating frequencies (with Transmitter); sensitivity on the selected frequency Frequency, Hz
    273
    491, 526
    982, 1024
    2000, 2048
    8440, 8928, 9820
    10000, 32768
    Sensitivity, min, μA/m*
    500
    300
    150
    50
    25
    5
    2. Rated values of operating frequencies (without Transmitter), Hz 50, 100, 300, 550 and 1450
    3. Dynamic range of input signals, dB, min 102
    4. Bandwidth for each operating frequency, max, Hz at the level of minus 3 dB at the level of minus 60 dB
    9 24
    5. Utility depth, m from 0,10 to 6,00
    6. Operating frequency current in the utility from 10 mA to 10,0 А
    7. Admissible error of the depth of single extended utility line, max ±{[4 + 0,3 h (h + 1)]%+0,1 m}, where h is a measured depth
    8. Admissible error of location of a single line at the depth of 2 m, min ±0,1
    9. Operating frequency voltage at the SENSOR jack input od 0,01 mV do 1,70 V
    10. Sensitivity at the SENSOR jack input at the signal-noise ratio of 6 dB, mV, min 0,05
    11. Limits of admissible main error of voltage measurement, max, % + dgt ± (3% m.v. + 3 digits)
    12. Input resistance at the SENSOR jack input, MOhm 1
    13. Overall dimensions, max, mm 700 x 300 x 140
    14. Continuous operation under the normal conditions with full battery, min, h 5
    15. Weight, max, kg 1,8

    * Rated for the “Broad peak” method. The signal-noise ratio is 6 dB.

  • Files

Standard accessories

Optional accessories

Slide 1 of 4

Basic in product

Slide 1 of 1